envelope-o at document map phones phone bars angle-left angle-right angle-up angle-down

PCA-40 high voltage capacitor charger

Applications

  • Capacitor charging for medical and industrial pulsed energy systems

Features

  • 4000 W nominal output power
  • Up to 2000 V output voltage
  • 200-240 V AC input
  • Meets all safety and EMC requirements for medical devices
  • Parallel operation
  • 3-year guarantee
pca-40-manual-en.pdf

Download user manual (1.2Mb)

PCA-40 capacitor charging power supply is high-frequency switch-mode converter, which transforms AC input to regulated high voltage DC output to charge capacitors. It was developed for the use in pumping systems of flashlamp-pumped solid-state laser devices.

PCA-40 is the most powerful modification of the PCA-series. Its maximum output power is 4000 W, so it is the best one for use in laser systems of high power like Long pulse Nd:YAG, Long pulse Alexandrite, 40-80 W Ho:YAG. The module is suitable to operate in parallel.

Several variants of PCA-40 are offered with different maximum output voltage level (Vmax), selected from 300 V to 1500 V at the moment of order. Higher output voltages are available on request. During operation an output voltage level can be chosen in working range (0 V - Vmax) by user, applying a DC voltage to the respective pin.

The actual output power depends on relation between actual operating voltage and Vmax. We would recommend to pick Vmax as close as possible to your operating voltage in order to utilize completely the output power of the device.

By default, modules are optimized for partial discharge applications. Complete discharge modifications are available on request.

Module is intended to be used for medical purposes, in medical laser equipment. All the device parameters meet the latest versions of medical standards requirements: IEC 60601-1:2005/AMD1:2012 for electrical safety and IEC 60601-1-2:2014 for EMC, so, PCA‑40 can be used in medical systems without any modifications, any additional filtration or additional means of protection.

High reliability and efficiency, easy to integrate OEM design and a lot of flexible parameters for customized solutions allows PCA-series to be used in wide variety of applications, both medical and industrial.

Input:

Voltage
200-240 V AC, 50/60 Hz
Current
< 24 A

Output:

Maximal output power
4000 W
Maximal output voltage
500/700/800/1000/1500 V standard modifications
other on request (up to 2000 V and higher)
Modifications
partial discharge modification,
complete discharge modification
Stability
< 0.5%
Ripple / hysteresis
< 0.5%
Efficiency
more than 85%

Safety:

PFC coefficient
> 0.98 (active)
Leakage current
< 300 uA
Safety standard
IEC 60601-1
EMC
IEC 60601-1-2
Protections
from open-circuit
from short-circuit
from overheating
from overvoltage

Environment:

Cooling
forced air (with built-in fan)
Operation temperature
+10 °C to +40 °C
Storage temperature
-20 °C to +60 °C
Humidity
0 ... 90%, non-condensing

Other:

Size (LxWxH)
296x153x132 mm
Weight
4.0 kg

Many parameters of the power supply can be taylored to your needs. Some of them are listed below:

  • output voltage incl. non-standard
  • optimization for complete discharge, partial discharge or other custom regime
  • reduced or bursted output power
  • interfaces (calibraton, logic)
  • type and location of connectors
  • cables
  • coverage case

Partial discharge modifications

Model *
Description
PCA-40-500V-PD
Maximal output voltage 500 V
Partial discharge modification
PCA-40-700V-PD
Maximal output voltage 700 V
Partial discharge modification
PCA-40-1000V-PD
Maximal output voltage 1000 V
Partial discharge modification
Model certificates 1, 2
PCA-40-1500V-PD
Maximal output voltage 1500 V
Partial discharge modification

(*) Other output voltages are available on request

Complete discharge modifications

Model *
Description
PCA-40-700V-CD
Maximal output voltage 700 V
Complete discharge modification
PCA-40-1000V-CD
Maximal output voltage 1000 V
Complete discharge modification
PCA-40-1300V-CD
Maximal output voltage 1300 V
Complete discharge modification
Model certificate 1
PCA-40-1500V-CD
Maximal output voltage 1500 V
Complete discharge modification

(*) Other output voltages are available on request

Example

PCA-40-1000V-PD
Maximal output voltage 1000 V
Partial discharge modification
pca-40-manual-en.pdf

Download user manual (1.2Mb)

pca-40-certificates.pdf

Download PCA-40 family Certificates of Conformity to IEC 60601-1, IEC 60601‑1‑2 and CE declaration of conformity (p/n - PCA‑40‑1000V‑PD, PCA‑40‑1300V‑СD)